Abstract
The radiocarbon dating equipment previously used (Dorn, Fairhall, Schell and Takashima, 1962) has been moved to a different location. During the resulting hiatus in our dating program we have constructed a new counter with a few novel features. It is similar in concept to the Houtermans-Oeschger counter (Houtermans and Oeschger, 1958), but two changes have been made in the design: (1) The inner counter, constructed of thin, metallized plastic foil of thickness 4.2 mg/cm2, is leak-tight relative to the outer counter. By means of two solenoid valves actuated by a simple differential mercury manometer the outer and inner counters can be filled separately with a pressure differential on the partition of less than 0.5 cm Hg. Thus all of the sample can be introduced into the inner counter while inert gas is fed into the outer counter. As the sample gas is not needed for anti-coincidence filling efficiency increases ca. 30%; (2) The metallic parts are made of commercially available high-purity nickel which is easier to procure than O.F.H.C. copper; nylon is used for the other parts. The outer counter appears to have a very low radioactivity, its α activity being 5 pulses/hr/100 cm2.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.