Abstract
A four-rod radio frequency quadruple (RFQ) accelerator is designed, manufactured, installed and commissioned for the Peking University Neutron Imaging Facility (PKUNIFTY). This 2699.6-mm-long RFQ accelerator with the mean aperture radius of 3.88mm is operating at 201.5 MHz in pulse mode. An inter-electrode voltage of 70 kV is needed to accelerate the injected 50 keV 40 mA D+ ions up to 2 MeV. We present the rf system, high rf power feeder design, lower rf measurements and higher rf power test. Especially, the rf commissioning was carried out with rf power up to ∼280 kW and duty factor of 4%. The measured x-ray spectrum shows that the rf inter-electrode voltage reaches 70.7 kV. It is found that the specific shunt impedance of the RFQ cavity reaches 52.7 kΩ·m.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.