Abstract
X-rays are by far most significant contributor to total population dose from man-made sources of radiation. Diagnostic reference levels provide frameworks to reduce variability. The aim of this study is to establish, for the first time, a baseline for national diagnostic reference levels in Serbia for the most common X-ray examination types. Dose estimates are based on measurements of kerma-area product and Entrance surface air kerma for at least ten patients for each examination type, in each of 16 randomly selected hospitals in Serbia. Mean, median and third quartile values of patient doses are reported. Results have shown wide variation of mean hospital doses. Entrance surface doses were compared with previously published diagnostic reference levels. Doses for all studied examination types except chest radiography were within European DRL. The reasons for dose variation are discussed. The findings emphasize the importance of regular patient dose measurement to ensure that patient doses are kept as low as reasonable achievable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.