Abstract

Based on the principle of Lorentz force induced acoustic vibration, radiation theory comparison between acoustic point and dipole sources was conducted for magnetoacoustic tomography with magnetic induction (MAT-MI). It is proved that each acoustic source of MAT-MI is produced by the divergence of the magnetically induced Lorentz force, and the detected acoustic pressure is the integral of all diffraction sources inside the object. Wave clusters are produced by abrupt pressure changes at conductivity boundaries, and only the configurations in terms of shape and size of phantom models can be reconstructed. However, different from point source, positive and negative pressures are generated by the radiation pattern of dipole sources. Reverse vibration phases of wave clusters in collected waveforms and opposite polarities of borderline stripes in reconstructed images are produced at conductivity boundaries, representing the direction of conductivity changes. The experimentally collected waveforms and reconstructed images of the aluminum foil cylinder and cylindrical saline gel phantom model agree well with simulated results. The favorable results prove the validity of the radiation theory of acoustic dipole source and provide basis for further investigation of conductivity reconstruction for MAT-MI.