Abstract

Scattered radiation makes up the majority of the stray radiation field around an X-ray unit. The scatter is linked to the amount of radiation incident on the patient. It can be estimated from quantities used to assess patient dose such as the kerma-area product, and factors have been established linking this to levels of scattered radiation for radiography and fluoroscopy. In radiography shielding against primary radiation is also needed, but in other modalities this is negligible, as the beam is intercepted by the image receptor. In the same way scatter from CT can be quantified in terms of dose-length product, but because of higher radiation levels, exposure to tertiary scatter from ceilings needs to be considered. Transmission requirements are determined from comparisons between calculated radiation levels and agreed dose criteria, taking into account the occupancy of adjacent areas. Thicknesses of shielding material required can then be calculated from simple equations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.