Abstract

Beam shaping in scattered and uniform scanned proton beam therapy (PBT) is made commonly by brass apertures. Due to proton interactions, these devices become radioactive and could pose safety issues and radiation hazards. Nearly 2,000 patient-specific devices per year are used at Indiana University Cyclotron Operations (IUCO) and IU Health Proton Therapy Center (IUHPTC); these devices require proper guidelines for disposal. IUCO practice has been to store these apertures for at least 4 mo to allow for safe transfer to recycling contractors. The devices require decay in two staged secure locations, including at least 4 mo in a separate building, at which point half are ready for disposal. At 6 mo, 20-30% of apertures require further storage. This process requires significant space and manpower and should be considered in the design process for new clinical facilities. More widespread adoption of pencil beam or spot scanning nozzles may obviate this issue, as apertures then will no longer be necessary.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.