Abstract

Tomotherapy, a novel radiotherapy technique, uses narrow fan beams for cancer patient treatment. Photon energy spectra for a rectangular 10 x 1 cm2 photon beam were analyzed in central axis and penumbra regions at depths of 3 to 10 cm in a water phantom. A 6 MV beam of a Varian 2100C/D Linear Accelerator was modeled using BEAM99 Monte Carlo calculations to simulate energy transport in a water phantom. Arrays of 4 x 2 mm2 scoring regions were arranged to cover the central axis and penumbra areas. Radiation quality factors were calculated based on dose-mean linear energy transfer. Although there appears to be a trend towards higher quality factor values in the penumbra area, this change is fairly small, at most 3% in penumbra region. We conclude that change in radiation quality is not likely to be an issue in a tomotherapeutic approach when 6 MV x rays are used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.