Abstract

Radiation induced electrical transients in strain gage, nickel temperature sensor, silicon temperature sensor, and thermocouple circuits have been studied in the Kukla Prompt Critical Pulsed Reactor environment. Induced signals, which could obscure real data, were observed in many strain gage circuits. It was determined that the observed signals were the result of a voltage appearing between the strain gages and ground when the reactor was pulsed. Circuit arrangements were devised and tested which reduced the magnitude of the induced signals to less than 1 mV. Nickel temperature sensors and spot welded thermocouples operated satisfactorily in the reactor environment. No adverse effects due to transient signals and no changes in calibration were observed. The one type of silicon temperature sensor tested suffered large and permanent increases in resistance when exposed to a reactor pulse.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.