Abstract
Here we report the results of an investigation into the effects of ionizing radiation on commercial off-the-shelf InGaAs and Si photodiodes. The photodiodes were exposed to 30, 52, and 98 MeV protons with fluences ranging from 108 - 5 × 1011 protons/cm2 at the Indiana University Cyclotron Facility. We tested the photodiodes for changes to their dark current and their relative responsivity as a function of wavelength. The Si photodiodes showed increasing damage to their responsivity with increasing fluence; the InGaAs photodiodes showed significantly increased dark current as the fluence increased. In addition, we monitored the absolute responsivity of the InGaAs photodiodes over their entire bandpass. Our measurements showed no evidence for broadband degradation or graying of the response at the fluences tested. All measurements in this investigation were made relative to detectors traceable to NIST standards.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.