Abstract
Perform a cadaveric experimental pilot study to measure and compare potential radiation exposure to an orthopedic surgeon from 2 different-generation mini C-arm models during a simulated orthopedic surgery. 16 radiation dosimeters. Mock surgery setups were constructed with a canine cadaver thoracic limb and 2 different-generation mini C-arm models. Four radiation dosimeters were placed near the mini C-arm to mimic common locations of radiation exposure during image acquisition. One mini C-arm was placed in auto technique mode, and 100 static images were acquired. The dosimeters were replaced, and a 5-minute-long dynamic image was acquired. The same protocols were repeated for the second mini C-arm. The dosimetry badges were then submitted for radiation exposure quantification. All but 1 dosimeter had radiation exposure levels below the detectable limits of the dosimeter. The dosimeter closest to the primary x-ray beam of 1 mini C-arm during dynamic image acquisition had a reading of 1 mrem. Intraoperative radiation exposure from the mini C-arm is low, specifically to areas not protected by lead and in close proximity to the primary x-ray beam. That being said, surgeons should always practice the principles of ALARA (ie, as low as reasonably achievable) to minimize radiation exposure in the workplace.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Veterinary Medical Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.