Abstract
The objectives of this study were to measure the entrance surface air kerma (ESAK) and the effective doses in pediatric patients undergoing micturating cystourethrography (MCU) procedures. The ESAK was determined using calibrated thermoluminescent dosimeters (TLD- GR 200 A) in 167 pediatric patients. The patient population was categorized into three groups according to age (0–5 years (group 1), 6–10 years (group 2) and 11–15 years (group 3)). The mean ESAK±SD and range (mGy) resulting from a MCU procedure was estimated to be 2.2±0.5 (0.8–9.2), 2.48±0.6 (0.9–8.1) and 3.90±0.6 (1.1–10.3) for group 1, 2 and 3, respectively. The mean effective dose was between 0.03 and 0.4 mSv per procedure for the pediatric population. Pediatric patients were exposed to avoidable radiation doses because no gonad shields were used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.