Abstract
Fluoroscopic guidance is commonly utilized during direct anterior total hip arthroplasty (DA THA). The purpose of this study was to measure patient and surgeon exposure utilizing this technique. Fifty-one consecutive patients who underwent primary DA THA by a single surgeon were prospectively studied. Fluoroscopic guidance was utilized according to an established protocol. Dose-area product (DAP) (Gy-cm2) and fluoroscopy time were recorded for each case. Surgeon exposure was recorded by a dosimeter. The median DAP was 0.716Gy-cm2 (range 0.251–1.81). Mean fluoroscopy time was 0.59minutes. Dosimeter results were 10 mrem for all procedures combined. DAP and fluoroscopy times were comparable to published values for other fluoroscopically guided hip procedures. This information may aid in setting reference dose levels for this procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.