Abstract
The aim of this study was to evaluate optimisation status during common computed tomography (CT) procedures by determining values of volume computed tomography dose index (CTDIvol) and dose-length product (DLP) per examination. Patient and exposure data were collected from the CT console during various CT procedures. The results show that variations in CTDIvol and DLP values were mainly because of differences in the techniques used. The 75th percentile values were set as the third quartile of the median CTDIvol or DLP values for all hospitals. These values of 40.9, 9.0, 9.4 and 16.2mGy for CTDIvol were determined for head, high-resolution chest, abdomen-pelvis and lumbar spine, respectively. The corresponding DLP values for the same sequence of CT procedures were 900, 360, 487 and 721mGy.cm, respectively. The updated results provide a basis for optimising the procedures of CT in this country.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.