Abstract

The aim of this study was to determine the cumulative effective doses (CED) from digital radiographic imaging in very low birth weight infants treated in a tertiary care neonatal intensive care unit (NICU). The CED for each infant was retrospectively calculated using a voxel-based model. The results were compared with previous studies applying conventional radiography. Two hundred and six preterm infants were included into this study. Neonates received a median of four radiographs (range: 1-68) and a CED of 50 µSv (4-883 µSv). Overall mean CED was lower than in previously published data applying conventional radiography. Factors contributing to a lower radiation dose per infant in our study were a lower number of radiographs and smaller field sizes per radiographic image. The number of conducted radiographs per patient and the employed field size had a higher impact on the CED than the applied radiographic technology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.