Abstract
Exposure to ionizing radiation in the operating room is governed by practical prevention and protection measures on the international, national and local levels. We evaluated the equivalent dose to the hand of an orthopedic surgeon over 13 months. An orthopedic surgeon wore a ring dosimeter on the ring finger of his right hand for all surgical procedures requiring intraoperative fluoroscopy between March 2014 and April 2015. Monthly doses were evaluated by the IRSN over the study period. The number and type of procedures were compiled as well as the type of fluoroscopy unit used. Four hundred procedures were performed during this period, including 182 with fluoroscopy. The equivalent cumulative dose at the hand was 4,75 mSv. No correlation was found with the type of procedure or type of fluoroscopy unit (conventional or mini C-arm). Equivalent doses were below the annual regulatory limit in France of 500 mSv. These results are consistent with those reported in the literature. However, recent studies have noted that both younger surgeons in training and more experienced surgeons must remember to use radiation protection measures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.