Abstract

The heating effect of normal laboratory illumination on the sensors of standard platinum resistance thermometers immersed in a standard water triple-point cell packed in crushed ice has been investigated. This study shows that the absorption by the platinum sensors of luminous and near-infrared radiation transmitted through the ice pack could easily amount to temperature errors as large as 0.000 5 °C. Any illumination error must be eliminated by making all measurements in the dark or under adequate radiation shielding.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.