Abstract

We estimated the dosimetry of [(18)F]fluoroacetate (FAC) with the method established by MIRD based on biodistribution data of rats. We selected some important organs and computed their residence time, their absorbed doses and effective dose with the (%ID(Organ)) (human) data using OLINDA/EXM 1.1 program. We observed the highest absorbed doses in the heart wall (0.025mGy/MBq) and the lowest in skin (0.0079mGy/MBq). The total mean absorbed doses and the effective doses were 0.011mGy/MBq and 0.014mSv/MBq, respectively. A 370-MBq injection of FAC leads to an estimated effective dose of 5.2mSv. The potential radiation risk associated with FAC/PET imaging is well within the accepted limits.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.