Abstract
The radiation dose to the lens during CT of the paranasal sinuses was measured in 20 patients. In 10 patients, a "standard" technique with axial and coronal sections was used. In the remaining 10 patients, overlapping axial sections for bidimensional reconstruction were obtained. Radiation dose was measured using thermoluminescent dosimeters. The mean dose was 22 mGy in patients who underwent "standard" CT and 42 mGy in those who underwent CT with overlapping sections. Dose was dependent on the total number of slices and of transorbital axial sections. Coronal sections did not statistically increase the radiation dose to the lenses. Although these doses were significantly below the critical dose for cataracts, CT with overlapping sections gives a considerably higher radiation dose. We recommend the bidimensional reconstruction technique only for selected cases (e.g. preoperative assessment of the roof of the ethmoid sinuses or in patients who can or should not maintain the hyperextended position required for coronal views.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.