Abstract

To compare the radiation dose and diagnostic ability of the 100-kVp protocol, based on the contrast noise ratio (CNR) index, during coronary artery bypass graft (CABG) vessels with those of the 120-kVp protocol. For the 120-kVp scans (150 patients), the targeted image level was set at 25 Hounsfield units (HU) (CNR120=iodine contrast/25 HU). For the 100-kVp scans (150 patients), the targeted noise level was set at 30 HU to obtain the same CNR as in the 120-kVp scans (i.e. using 1.2-fold higher iodine contrast, CNR100=1.2×iodine contrast/(1.2×25 HU)=CNR120). We compared the CNRs, radiation doses, detection of CABG vessels and visualisation scores of the scans acquired at 120 and 100 kVp, respectively. At the same CNR, the 100-kVp protocol may help reduce the radiation dose by ⁓30% compared with the 120-kVp protocol, without degradation of diagnostic ability during CABG.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.