Abstract

IntroductionRadiation dose to patients undergoing invasive coronary angiography (ICA) is relatively high. Guidelines suggest that a local benchmark or diagnostic reference level (DRL) be established for these procedures. This study sought to create a DRL for ICA procedures in Queensland public hospitals.MethodsData were collected for all Cardiac Catheter Laboratories in Queensland public hospitals. Data were collected for diagnostic coronary angiography (CA) and single-vessel percutaneous intervention (PCI) procedures. Dose area product (PKA), skin surface entrance dose (KAR), fluoroscopy time (FT), and patient height and weight were collected for 3 months. The DRL was set from the 75th percentile of the PKA.Results2590 patients were included in the CA group where the median FT was 3.5 min (inter-quartile range = 2.3–6.1). Median KAR = 581 mGy (374–876). Median PKA = 3908 uGym2 (2489–5865) DRL = 5865 uGym2. 947 patients were included in the PCI group where median FT was 11.2 min (7.7–17.4). Median KAR = 1501 mGy (928–2224). Median PKA = 8736 uGym2 (5449–12,900) DRL = 12,900 uGym2.ConclusionThis study established a benchmark for radiation dose for diagnostic and interventional coronary angiography in Queensland public facilities.