Abstract
Doses for a range of examinations and views using digital X-ray equipment with full-body linear slit scanning capabilities (Statscan) have been compared with those from other published studies. Entrance doses (free-in-air) were measured using a dosimeter, and effective doses were generated using a Monte Carlo simulator. Doses delivered by the linear slit scanning system were significantly lower than those from conventional X-ray equipment. Effective doses were between 9 and 75% of the United Nations Scientific Committee Report on the Effects of Ionising Radiation doses for standard examinations. This dose reduction can be explained by the properties of linear slit scanning technology, including low scatter, beam geometry, the use of a digital detector and the use of higher than usual tube voltages.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.