Abstract

The dose from ionizing radiation exposure can be interpolated from a calibration curve fit to the frequency of dicentric chromosomes (DCs) at multiple doses. As DC counts are manually determined, there is an acute need for accurate, fully automated biodosimetry calibration curve generation and analysis of exposed samples. Software, the Automated Dicentric Chromosome Identifier (ADCI), is presented which detects and discriminates DCs from monocentric chromosomes, computes biodosimetry calibration curves and estimates radiation dose. Images of metaphase cells from samples, exposed at 1.4-3.4 Gy, that had been manually scored by two reference laboratories were reanalyzed with ADCI. This resulted in estimated exposures within 0.4-1.1 Gy of the physical dose. Therefore, ADCI can determine radiation dose with accuracies comparable to standard triage biodosimetry. Calibration curves were generated from metaphase images in ~10 h, and dose estimations required ~0.8 h per 500 image sample. Running multiple instances of ADCI may be an effective response to a mass casualty radiation event.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.