Abstract

AbstractBackgroundIn radiation therapy, to spare normal surrounding tissues, either Multileaf Collimators or Cerrobend blocks are used.PurposeThe current study focuses on the relative dose distribution under the areas protected by Cerrobend blocks.Materials and methodsA dual-energy linear accelerator and a Cobalt-60 machine were utilised as radiation sources. Several blocks were designed using commercially available materials to shield radiation fields. The relative dose distribution was then evaluated using extended dose range 2 films.ResultsResults showed that the dose distribution under protected areas depends on several parameters including the width and height of protecting blocks, incident photon beam energy, radiation field size and source to surface distance. An increase in Cerrobend block height from 80 to 95 mm significantly decreases the dose at the protected areas.ConclusionAn increase in the block width and photon energy decreases the relative dose deposition at the protected area. However, electron and neutron contaminations should also be taken into consideration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.