Abstract
Hysterosalpingography using a digital and conventional image intensification technique was carried out on two groups of 32 patients as part of investigation for infertility. Radiation dose was recorded as surface dose. In 51.7% of digital images and 54.8% of conventional images the findings were normal. Surface dose with conventional technique was 442 +/- 144 cGy/cm2, significantly higher than 207 +/- 113 cGy/cm2 with digital image intensification. The number of exposures was also significantly higher when using the conventional technique. The digital images showed no lack of diagnostic detail when compared with the conventional 100 mm films.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.