Abstract

Two high-purity germanium coaxial detectors, having opposite electrode configurations from one another, but fabricated from the same germanium crystal, were irradiated simultaneously with fast neutrons from an unmoderated 252Cf source. Both detectors were 42 mm diam. The detector having the conventional electrode configuration was about 28 times more sensitive to radiation damage than was the detector having the p+ contact on the coaxial periphery. These results prove that germanium coaxial detectors having the conventional electrode configuration should not be used in any situation subject to significant radiation damage. This conclusion was anticipated because the defects produced by neutron and proton irradiation of germanium act predominantly as hole traps.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.