Abstract

The BEAMnrc code was used for the simulation of the Theratron Equinox-80 telecobalt machine. The phase space of radiation beam was generated at treatment distance of 80 cm for various field sizes. The phase spaces in air were analyzed by BEAMdp data processing program. The electron energy fluence with respect to photon energy was 0.09% and 0.34% for field size of 05 × 05 and 35 × 35 cm2 respectively and it was maximum at the central axis which gradually decreases beyond this. The profiles for photon fluence were in symmetry for all the fields. The full width at half maximum of profiles in photon energy fluence shows good agreement with the field size. The photon energy fluence was flat till the field size of 27 cm2, after which it decreases gradually till the edge in larger field sizes. The air-kerma output factor from the simulation was in good agreement with measured value. We analyzed the dose data scored in the voxels in a large water phantom by simulation using dosxyznrc code. The percentage depth dose for all field sizes was in good agreement with the BJR supplement 25 and the data supplied by the manufacturer of machine. Significant deviation of about 20% in isodose line near the edge of the profile was observed for 35 × 35 cm2 field size. The penumbra widths of all field sizes were comparable except for 35 × 35 cm2, which has a penumbra width of 4.1 cm at 10 cm depth. The significant under dose near the edge as compared to central axis for larger field sizes may be the indication for its careful use in treatment.

Highlights

  • There is a long history of use of telecobalt unit in treatment of cancer

  • The percentage depth dose for all field sizes was in good agreement with the BJR supplement 25 and the data supplied by the manufacturer of machine

  • The present study performs the simulation of Theratron Equinox-80 telecobalt machine using BEAMnrc code for beam analysis and dose calculation in water phantom

Read more

Summary

Introduction

There is a long history of use of telecobalt unit in treatment of cancer. The 3-D conformal radiation therapy is performed with the availability of the treatment planning system and radiation field analyzer. The new treatment modalities such as intensity modulated radiotherapy and image guided radiotherapy are facilitated in the telecobalt unit [1] [2] [3]. The dosimetry and treatment planning in the telecobalt unit is straightforward as compared to the linear accelerator. The radiation field analyzer is used for the measurement of the depth-dose data and dose profiles for the purpose of input data in treatment planning system. Several authors have performed the simulation of various telecobalt units for the study of dosimetry parameters [5]-[12]

Methods
Results
Conclusion