Abstract
A simple technique has been developed to image intense pulsed proton beams with radiachromic film and to measure their spatial distribution. The optical density (OD) of 50 μm thick film, sufficient to stop 1.5 MeV protons, is measured to infer the incident beam fluence. The OD increases nearly linearly with fluence up to 0.25 cal/cm2. This film is used to measure the uniformity of large-area beams as well as the detector in a multiple-pinhole camera to determine the source uniformity and divergence of applied-magnetic field and pinched-beam ion diodes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.