Abstract
Background.Using radar for non-contact measuring human vital signs has garnered significant attention due to its undeniable benefits. However, achieving reasonably good accuracy in contactless measurement senarios is still a technical challenge.Materials and methods.The proposed method includes two stages. The first stage involves the process of datasegmentation and signal channel selection. In the next phase, the raw radar signal from the chosen channel is subjected to modified Pan-Tompkins.Results.The experimental findings from twelve individuals demonstrated a strong agreement between the contactless radar and contact electrocardiography (ECG) devices for heart rate measurement, with correlation coefficient of 98.74 percentage; and the 95% limits of agreement obtained by radar and those obtained by ECG were 2.4 beats per minute.Conclusion.The results showed high agreement between heart rate calculated by radar signals and heart rate by electrocardiograph. This research paves the way for future applications using non-contact sensors to support and potentially replace contact sensors in healthcare.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.