Abstract

Characterization of digital X-ray imaging devices is very important because it can be used to measure and compare the performance of detectors used in Diagnostic Radiology. This characterization is usually made through the calculation of Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE). These parameters, especially the DQE, are very important because they quantify the effect of spatial resolution, contrast and noise on Radiographic image quality (IQ). The IEC 62220-1-1:2015 International Standard provides comprehensive guidelines how to capture and analyze X-ray images to characterize digital X-ray detectors. A novel, fast and free MATLAB-based software was developed, named RAD_IQ, to calculate the Signal Transfer Property (STP), perform Noise Component Analysis (NCA), and calculate the parameters MTF, NPS & DQE of X-ray detectors based on the novel IEC 62220-1-1:2015 International Standard for General Radiography and IEC 62220-1-1:2007 for Digital Mammography. Our results were validated against well-established software products used for quantitative image analysis of digital X-ray detectors. The calculated parameters were within 5% difference compared to available software products. The conclusion of our study was that RAD_IQ can be easily used from Medical Physicists, Biomedical Engineers and researchers without any programming experience to characterize the performance of digital X-ray detectors used in Diagnostic Radiology.