Abstract

<h2>Abstract</h2> The case of a 17-year-old female who had been implanted a dual-chamber DDD pacemaker because of third-degree atrioventricular block is reported. There is a history of continued dizziness and even occasional syncopes. At heart rates of 111/min to 124/min, 24-h Holter electrocardiography revealed isolated missing ventricular beats in an otherwise continuous atrially sensed and triggered, ventricularly paced rhythm. Differential diagnoses of a putative pacemaker dysfunction are presented, comprising 2:1-block at maximum programmed heart rate, intermittent lead fracture, anti-pacemaker-mediated tachycardia algorithm, ventricular oversensing, P wave signal undersensing, and atrial oversensing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.