Abstract

The "Internal Fixator" for stabilizing the dorso-lumbar spine represents an effective and meanwhile established device to neutralize flexion-bending and sagittal shearing forces. Because of the possible movement of the Schanz' screws in the bone and within the connection clamps, this device does not withstand torsional and frontal plane shearing forces after postinjury or postlaminectomy instability. Therefore, a cross-link device, adaptable to the "Internal Fixator" has been developed. Biomechanical testing against torsion and frontal plane bending moments have shown very low stability resulting in lateral displacement of the fixation device without cross-linking. The diagonal bracing with two wire cerclages does not significantly increase the stability. The recently developed cantilever cross-link device completes the two longitudinal bars to a frame construction and increases therefore the torsional and frontal plane stability for over 70%. This supplementary device is easy to apply to the "Internal Fixator" and has been successful in clinical practise.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.