Abstract

The external standard source of liquid scintillation counter, 226Ra, was exchanged to 152Eu. Quench correction curves by irradiating the radiation to quenched standard samples for a single labeled 3H or 14C, or dual labeled 3H and 14C, were plotted by the computer software that controls the liquid scintillation counting system. Using the correction curve, the radioactivity of quenched standard sample for making correction curves, was determined again quantitatively. The most of corrected values were within ±5% error of the nominal radioactivity of the standard sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.