Abstract

A nitric oxide (NO) sensor employing a thermoelectrically cooled, continuous-wave, distributed feedback quantum cascade laser operating at 5.47 μm (1828 cm-1) and off-axis integrated cavity output spectroscopy was used to measure NO concentrations in exhaled breath. A minimum measurable concentration (3σ) of 3.6 parts-per-billion by volume (ppbv) of NO with a data-acquisition time of 4 s was demonstrated. Five prepared gas mixtures and 15 exhaled breath samples were measured with both the NO sensor and for intercomparison with a chemiluminescence-based NO analyzer and were found to be in agreement within 0.6 ppbv. Exhaled NO flow-independent parameters, which may provide diagnostic and therapeutic information in respiratory diseases where single-breath measurements are equivocal, were estimated from end-tidal NO concentration measurements collected at various flow rates. The results of this work indicate that a laser-based exhaled NO sensor can be used to measure exhaled nitric oxide at a range of exhalation flow rates to determine flow-independent parameters in human clinical trials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.