Abstract

A compact, fast-response, infrared spectrometer using pulsed quantum cascade (QC) lasers has been applied to the analysis of gases in mainstream and sidestream cigarette smoke. QC lasers have many advantages over the traditional lead salt tunable diode lasers, including near-room-temperature operation with thermoelectric cooling and single mode operation with improved long-term stability. The instrument uses a 36-m, 0.3-L multiple pass absorption gas cell to obtain a time response of 0.1 s for the mainstream system and 0.4 s for the sidestream system. A precision of 10-4 absorbance units in 1 s is achieved using pulse normalization to reduce pulse-to-pulse intensity variations. Laser line widths (hwhm) of 0.006 cm-1 at 967 and 0.010 cm-1 at 1900 cm-1 are obtained in pulsed operation. With this instrument, we have measured the concentrations of ammonia and ethylene using a QC laser at 967 cm-1 and nitric oxide using a laser at 1900 cm-1 in both mainstream and sidestream smoke from two types of cigarettes. ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.