Abstract

The accuracy of two calculation algorithms of the Eclipse 8.9 treatment planning system (TPS)--the anisotropic analytic algorithm (AAA) and pencil-beam convolution (PBC)--in modeling the enhanced dynamic wedge (EDW) was investigated. Measurements were carried out for 6 and 18 MV photon beams using a 2D ionization chamber array. Accuracy of the TPS was evaluated using a gamma index analysis with the following acceptance criteria for dose differences (DD) and distance to agreement (DTA): 3%/3 mm and 2%/2 mm. The TPS models the dose distribution accurately except for 20×20 cm(2) field size, 60 (°) and 45 (°) wedge angles using PBC at 6 MV photon energy. For these latter fields, the pass rate and the mean value of gamma were less than 90% and more than 0.5, respectively at the (3%/3 mm) acceptance criteria. In addition, an accuracy level of (2%/2 mm) was achieved using AAA with better agreement for 18 MV photon energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.