Abstract
This paper describes a soft and a hard tensile stage for the high voltage electron microscope. Both stages can be combined with a top entry double tilting stage. The devices are driven by thermal expansion elements operating against water cooled parts. The drive mechanism shows smooth action with a relatively low response time and good long-time stability. The stages are equipped with strain gauge bridges for force and elongation measurements. The soft stage has a maximum load of 13 g, and a hard one reaches 1.5 kg.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.