The rate of complete resection of glioma has improved with the introduction of 5-aminolevulinic acid-induced protoporphyrin IX (PpIX) fluorescence image guidance. Surgical outcomes are further enhanced when the fluorescence signal is decoupled from the intrinsic tissue optical absorption and scattering obtained from diffuse reflectance measurements, yielding the absolute PpIX concentration, [PpIX]. Spatial frequency domain imaging was used previously to measure [PpIX] in near-surface tumors under blue fluorescence excitation. Here, we extend this to subsurface [PpIX] fluorescence under red-light excitation. The decay rate of the modulation amplitude of the fluorescence signal was used to calculate the PpIX depth, which was then applied in a forward diffusion model to estimate [PpIX] at depth. For brain-like optical properties in phantoms with PpIX fluorescent inclusions, the depth can be recovered up to depths of 9.5 mm ± 0.4 mm, with [PpIX] ranging from 5 to 15 μg/mL within an average deviation of 15% from the true [PpIX] value.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.