Abstract

The authors describe the prospective use of a new hand-held point-and-shoot pupillometer (NeurOptics) to assess pupil function quantitatively. Repetitive measurements were made in 90 pediatric participants ranging in age from 1 to 18 years, providing a total of 100 measurements under ambient light conditions. The participants consisted of 45 patients without known intracranial or ophthalmological pathological conditions as well as 45 volunteers in the outpatient setting. Quantitative pupil measurements were reliably replicated in the study participants. The mean resting pupil aperture was 4.11 mm and the minimal diameter after stimulation was 2.65 mm, resulting in a 36% change in pupil size. The mean constriction velocity was 2.34 mm/second, with a mean dilation velocity of 2.2 mm/second. Pupil symmetry was impressive in the entire cohort.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.