Abstract

Track-etch detectors made of cellulose nitrate (LR 115, Kodak Pathe) and polycarbonate (CR 39, Pershore Mouldings Ltd.) were compared regarding sensitivity and background when used as detectors for boron determination in biological samples. Measurements were made on two kinds of sample, cryosectioned biological tissue, and liquid samples deposited directly on the detector surface as microdroplets. The CR 39 films were pretreated or washed before irradiation. When cryosectioned tissue was used, measurements were made with and without the inclusion of Mylar foils between the samples and the detectors. Foil thicknesses used were 2 mu m in the case of LR 115 and 2, 4, and 6 mu m in the case of CR 39. All samples were irradiated with a thermal-neutron fluence of 5*1012 neutrons cm-2 at the thermal-neutron facility in Studsvik, Sweden. The use of a Mylar foil generally suppressed the background tracks relative to the tracks from the 10B disintegration. No difference in resolution between CR 39 and LR 115 was observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.