Abstract

A 20 MHz pulsed Doppler ultrasound system was used to obtain steady flow velocity profiles in cylindrical tubes. A deconvolution technique was employed to improve the spatial resolution and the corrected results were compared with Poiseuille's theory. Frequency domain windowing was used to suppress noise in the deconvolution procedure, using a method which reduced distortions at the window boundaries. Wall shear rate was calculated from the measured velocity profiles both before and after deconvolution, and compared with theory to quantify the accuracy of the technique. The results illustrate the significant improvement provided by deconvolution, with the average error in measured wall shear rate reducing from 46 to 12% after applying the technique. Some of the discrepancy is in part attributed to the method used to determine wall shear rate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.