Abstract

In the present study, forward degenerate four-wave mixing (FDFWM) is employed to measure rubidium (Rb) isotope ratio quantitatively with excellent reproducibility regardless of laser intensity turbulence. The effects of power saturation, atomic absorption and Doppler broadening on measurements are investigated. The results indicate that optimized pump power for isotope ratio measurement is 5–37.3mW and the optimized atomic number density is around 1014m−3 for the sample cell with a length of 50mm. In addition, the signal of FDFWM is almost independent on Doppler effect in isotope ratio determination. The measured 85Rb and 87Rb ratios in Rb vapor cell and RbCl are 2.590±0.001 and 2.649±0.002 respectively, which are in good agreement with the values obtained by absorption spectroscopy and mass spectrometry. A low detection limit as 18.41fg/mL is achieved. The information obtained from this work suggests that FDFWM is a reliable laser spectroscopic technique in quantitative isotope trace analysis for various applications such as geological and archeological dating.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.