Abstract
Femtosecond pulse characterization and compensation using multiphoton intrapulse interference phase scan (MIIPS) [Opt. Lett.29, 775 (2004)] was rigorously tested. MIIPS was found to have 3 mrad precision within the 90 nm bandwidth of the pulses. Group-velocity dispersion measurements of glass and quartz provided independent accuracy tests. Phase distortions from high-numerical-aperture objectives were measured and corrected using MIIPS, an important requirement for reproducible two-photon microscopy. Phase compensation greatly improved the pulse-shaping results through a more accurate delivery of continuous and binary phase functions to the sample. MIIPS measurements were possible through the scattering of biological tissue, a consideration for biomedical imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.