Abstract

This paper describes the framework of quality by design applied to the development, optimization and validation of a sensitive capillary electrophoresis-sodium dodecyl sulfate (CE-SDS) assay for monitoring impurities that potentially impact drug efficacy or patient safety produced in the manufacture of therapeutic MAb products. Drug substance or drug product samples are derivatized with fluorogenic 3-(2-furoyl)quinoline-2-carboxaldehyde and nucleophilic cyanide before separation by CE-SDS coupled to LIF detection. Three design-of-experiments enabled critical labeling parameters to meet method requirements for detecting minor impurities while building precision and robustness into the assay during development. The screening design predicted optimal conditions to control labeling artifacts while two full factorial designs demonstrated method robustness through control of temperature and cyanide parameters within the normal operating range. Subsequent validation according to the guidelines of the International Committee of Harmonization showed the CE-SDS/LIF assay was specific, accurate, and precise (RSD ≤ 0.8%) for relative peak distribution and linear (R > 0.997) between the range of 0.5-1.5 mg/mL with LOD and LOQ of 10 ng/mL and 35 ng/mL, respectively. Validation confirmed the system suitability criteria used as a level of control to ensure reliable method performance.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call