Abstract
The purpose of this study was to evaluate global quantitation of cardiac uptake on I-123 mIBG SPECT. The study included a pilot group of 67 subjects and a validation group of 1,051 subjects. SPECT images were reconstructed by filtered backprojection, ordered subsets expectation maximization, and deconvolution of septal penetration, respectively. SPECT heart-to-mediastinum ratio (H/M) was calculated by comparing the mean counts between heart and mediastinum volumes of interest drawn on transaxial images. Receiver operating characteristic (ROC) analysis was used to assess the capability of each SPECT method to differentiate the heart disease subjects from controls in comparison with that of the planar H/M. In the validation group, the areas under the ROC curves were not significantly different between the SPECT and planar H/M. Order subsets expectation maximization had significantly larger area under the ROC curve than the other two SPECT methods. H/M obtained from I-123 mIBG SPECT was equivalent to the planar H/M for differentiating between subjects with normal and abnormal mIBG uptake. Global quantification of cardiac I-123 mIBG SPECT may represent a viable alternative to the planar H/M.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.