Volatile compounds in various solid samples have been analyzed quantitatively by equilibrium headspace gas chromatography with the technique of multiple headspace extraction (MHE). Solid samples can represent either a partition system, such as certain polymers, or an adsorption system, particularly when present as porous material. It has been found that the humidity of the sample has a strong influence on the equilibration in an adsorption system with a hydrophilic matrix and that the addition of water increases the volatility and speeds up the equilibration time. The influence of various important parameters, such as temperature, time, size and structure of the solid samples on the equilibration is demonstrated on the examples of the quantitative determination of vinyl chloride monomer in PVC, styrene in polystyrene, ethylene in polyethylene, ethylene oxide in a contact lens, halogenated hydrocarbons in coffee and residual solvents in drugs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.