Abstract

BackgroundAtrial and ventricular arrhythmias significantly contribute to morbidity and mortality of patients with cardiac disease. Ablation of these arrhythmias has shown to improve clinical outcomes, yet targeted ablation strategies rely on proper mapping capabilities. In the present study, we compare different modes of high-resolution mapping in clinically relevant arrhythmias using HD grid. Methods and resultsUsing the Advisor™ HD Grid Mapping Catheter in either the standard, the wave (bipolar along spline and bipolar orthogonal) or the wave diagonal setting, low-voltage areas were determined. Low-voltage was defined as local electrograms with an amplitude <0.5 mV (bipolar; atria/ventricle) or <4 mV (unipolar; ventricle). Ultra high-density mapping in 47 patients with ventricular tachycardia, ventricular premature beats, atrial fibrillation and atrial tachycardia provided reliable information for the understanding of the arrhythmia mechanism resulting in safe ablation procedures. Regions of low voltage were significantly decreased by 14 ± 2% and 31 ± 3% with wave and wave diagonal settings as compared to standard settings, respectively. ConclusionSubstrate mapping and risk stratification relies on proper low voltage discrimination. Even though the Advisor™ HD Grid Mapping Catheter was safely used in all cases, the extent of low voltage areas was mapping-mode dependent.