A simple method is proposed in this work for quantitative evaluation of the quality of the threshold adjustment and the flat-field correction of Hybrid Photon Counting pixel (HPC) detectors. This approach is based on the Photon Transfer Curve (PTC) corresponding to the measurement of the standard deviation of the signal in flat field images. Fixed pattern noise (FPN), easily identifiable in the curve, is linked to the residual threshold dispersion, sensor inhomogeneity and the remnant errors in flat fielding techniques. The analytical expression of the signal to noise ratio curve is developed for HPC and successfully used as a fit function applied to experimental data obtained with the XPAD detector. The quantitative evaluation of the FPN, described by the photon response non-uniformity (PRNU), is measured for different configurations (threshold adjustment method and flat fielding technique) and is demonstrated to be used in order to evaluate the best setting for having the best image quality from a commercial or a R&D detector.