Abstract

Application of quantitative X-ray imaging to frozen hydrated tissue sections has presented a number of major problems including lack of a suitable algorithm which could deal effectively with mass loss due to radiation damage, problems of low characteristic X-ray signal to background ratios, and provide a means of analysis of the same location in both hydrated and dried states. This paper presents details of the application of our algorithm for analysis of frozen hydrated, then dried cryosections applied to quantitative X-ray imaging, which provides relatively high precision quantitative measurement of elemental content (related to both wet and dry weight) and water content of each pixel. This algorithm largely circumvents many of the problems of analysis of frozen hydrated tissue sections. Our algorithm for X-ray imaging obtains reasonably precise quantitative measurements coupled with morphological information by trading speed and image resolution.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call