Abstract

A high performance liquid chromatographic method for the determination of hippuric and m-(p-)methylhippuric acids in urine, metabolites of toluene and m-(p-)xylene, is described. A stainless-steel column packed with octadecyl silanized silica gel was used and the mobile phase was a mixed solution of methanol/water/acetic acid (20/80/0.2). The method is simple and specific. Urine can be analyzed directly, without solvent extraction or pretreatment. The method has a lower detection limit of 0.2 μg on column. All the analysis and quantitative determination can be performed within about 30 min. for samples containing m-(p-)methylhippuric acid.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.