Abstract

We have developed a simple, rapid and highly sensitive method for determining a plasma or cerebrospinal fluid (CSF) concentrations of individual teicoplanin components using reversed-phase high-performance liquid chromatography followed by electrochemical detection. A linear relationship was observed between concentrations and peak heights for the teicoplanin concentration range of 0.025–10 μg/mL. The correlation coefficients of all standard curves were greater than or equal to 0.999. The limit of detection for the major component of teicoplanin was 1.0 ng/mL (signal/noise ratio >3). Daily fluctuations of standard curves ( n = 5) were small, with coefficients of variation of 3.3%. The intra-assay precision was 5.9% ( n = 5). Inter-assay precision ranged from 2.6 to 6.8%. The method described here is suitable for clinical monitoring of teicoplanin levels in plasma or CSF level and for use in studies involving pharmacokinetics of individual teicoplanin component.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.